Pressure Steam Sterilization Chemical Indicator Card

This instruction card is suitable for pre vacuum pressure steam sterilization as a chemical monitoring of sterilization effectiveness.

Scope Of Use

Suitable for pre vacuum pressure steam sterilization as a chemical monitoring of sterilization effectiveness.

Usage Method

Place this instruction card in the center of the sterilization package or in a location where steam is difficult to reach. After sterilization, remove the indicator card and observe its color change to determine if it meets the sterilization qualification requirements.

Color Change Instructions

If the indicator on the card changes from beige to black within the temperature range of 132 ° C-135 ° C for pre realistic pressure steam, it indicates "sterilization qualified". Otherwise, it is considered unqualified. For packages that are "unqualified", they cannot be used and should be re sterilized. The cause should be promptly investigated until the monitoring results meet the requirements.

Matters Needing Attention

1. This product should not come into direct contact with water, and the indicator card should not be placed directly on surfaces such as metal or glass materials that are prone to condensation.

2. Suitable for chemical monitoring of steam sterilization at 132 ° C-135 ° C, not for monitoring of dry heat and chemical gas sterilization.

3. Not applicable for testing the effectiveness of steam sterilization at 121 ° C.

Storage Method

1. Store at room temperature of around 20 ℃ and relative humidity of 50%. Avoid exposure to sunlight, fluorescent lights, and ultraviolet radiation.

2. Do not come into contact with acidic or alkaline substances. Moisture proof, store in a cool and shady place.

Term Of Validity

The storage validity period is 12 months.